Extech PH110

Extech PH110 Waterproof ExStik pH Meter Instruction Manual

Brand: Extech | Model: PH110

1. Rhagymadrodd

The Extech PH110 Waterproof ExStik pH Meter is designed for accurate pH measurements in various substances. This device features a refillable pH electrode, allowing for extended use and consistent accuracy by replenishing the reference solution. Its robust, waterproof design ensures reliability in diverse environments.

2. Nodweddion Allweddol

  • Refillable pH electrode for extended service life.
  • Analog bargraph display on LCD for trend indication.
  • Waterproof construction to IP57 standard.
  • Automatic 1, 2, or 3 point calibration with buffer solution recognition.
  • Flat Surface Electrode capable of measuring pH in liquids, semi-solids, and solids.

3. Cynnwys Pecyn

The Extech PH110 Waterproof ExStik pH Meter package includes:

  • Extech PH110 pH Meter with refillable pH electrode
  • 1.5mL bottle of KCI (Potassium Chloride) solution
  • Two reference junctions
  • Two electrode removal tools
  • Cap amddiffynnol
  • Sample cwpan
  • Four 3V CR-2032 batteries (pre-installed or included separately)
  • 48-inch (1.2m) neck strap

4. Gosod

4.1 Gosod Batri

  1. Ensure the meter is off.
  2. Lleolwch y compartment batri.
  3. Insert four 3V CR-2032 Lithium Metal batteries, observing correct polarity.
  4. Caewch y compartment batri yn ddiogel.

4.2 Electrode Preparation

  1. Tynnwch y cap amddiffynnol o'r electrod.
  2. Rinse the electrode tip with distilled water.
  3. If the electrode appears dry, soak it in electrode storage solution or pH 7 buffer for at least 30 minutes before first use.

5. Gweithrediad

5.1 Graddnodi

Calibration is essential for accurate measurements. The PH110 supports 1, 2, or 3 point calibration and automatically recognizes buffer solutions.

  1. Turn on the meter by pressing the YMLAEN / I FFWRDD botwm.
  2. Prepare fresh pH buffer solutions (e.g., pH 4, 7, and 10). These are sold separately.
  3. Pwyswch a dal y CAL button until 'CAL' appears on the display.
  4. Immerse the electrode into the first buffer solution (e.g., pH 7). The meter will automatically recognize the buffer and calibrate. Wait for the reading to stabilize.
  5. For 2 or 3 point calibration, rinse the electrode with distilled water and immerse it into the next buffer solution (e.g., pH 4, then pH 10). The meter will continue to calibrate automatically.
  6. Ensure the display shows 'CAL' during the calibration process.

5.2 Mesur

The flat surface electrode allows for versatile measurements.

  1. After calibration, rinse the electrode with distilled water.
  2. Trochwch yr electrod yn y sample to be measured. For liquids, ensure the electrode tip is fully submerged. For semi-solids or solids, ensure good contact with the flat surface electrode.
  3. Allow the reading to stabilize on the display. The analog bargraph will indicate trends in acidity or alkalinity.
  4. To switch between pH and temperature display modes, press the MODD botwm.
  5. After measurement, rinse the electrode thoroughly with distilled water and replace the protective cap.
Extech PH110 Waterproof ExStik pH Meter displaying a pH reading.

Ffigur 1: The Extech PH110 pH meter, a handheld device with a digital display showing pH and temperature readings. It features three control buttons: MODE, CAL, and ON/OFF.

6. cynnal

6.1 Electrode Refilling

The refillable electrode design allows for prolonged accuracy and lifespan.

  1. When the reference solution depletes or accuracy declines, use the provided removal tools to access the reference junctions.
  2. Carefully refill the electrode with the 1.5mL KCI solution.
  3. Ensure the reference junctions are properly re-seated after refilling.
Hands refilling the Extech PH110 pH meter electrode with filling solution.

Ffigur 2: A close-up view of hands refilling the Extech PH110 pH meter's electrode module with a specialized electrode filling solution from a small bottle. This process ensures the meter's continued accuracy and extends electrode life.

6.2 Glanhau a Storio

  • Always rinse the electrode with distilled water after each use to prevent contamination.
  • The flat surface electrode is designed for easy cleaning. Gently wipe with a soft cloth if necessary.
  • Store the electrode with its protective cap filled with electrode storage solution (or pH 7 buffer if storage solution is unavailable) to keep the sensor hydrated. Do not store dry.

7. Datrys Problemau

If the meter exhibits inaccurate readings or operational issues, consider the following:

  • Darlleniadau Anghywir: Recalibrate the meter using fresh buffer solutions. Ensure the electrode is clean and properly hydrated. Check if the reference solution in the electrode needs refilling.
  • Dim Problemau Arddangos/Pŵer: Verify that the batteries are correctly installed and have sufficient charge. Replace batteries if necessary.
  • Ymateb Araf: The electrode might be dry or contaminated. Clean and rehydrate the electrode, then recalibrate.

8. manylebau

ManylebManylyn
GwneuthurwrExtech
Rhif RhanPH110
Pwysau Eitem3.87 owns
Dimensiynau Cynnyrch6.8 x 1.6 x 1.4 modfedd
Rhif Model yr EitemPH110
Batris4 Lithium Metal batteries (CR-2032) required (included)
Ffynhonnell PwerWedi'i Bweru â Batri
Graddfa dal dwrIP57 safonol
Calibradu1, 2, or 3 point automatic buffer recognition
Math electrodRefillable Flat Surface Electrode

9. Gwybodaeth Gwarant

Extech provides a full one-year warranty on parts and workmanship for the PH110 pH Meter, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modifications.

10. Cefnogaeth

For further assistance, technical support, or to purchase accessories and replacement parts, please refer to the official Extech Instruments websafle neu cysylltwch â'u hadran gwasanaeth cwsmeriaid. Gellir dod o hyd i wybodaeth gyswllt fel arfer ar becynnu'r gwneuthurwr neu websafle.

Dogfennau Cysylltiedig - PH110

Cynview Extech ExStik™ Waterproof pH Meters PH100 & PH110 User Guide
Comprehensive user guide for the Extech ExStik™ Waterproof pH Meters, models PH100 and PH110. Learn about operation, calibration, maintenance, and specifications.
Cynview Llawlyfr Defnyddiwr Mesuryddion pH Gwrth-ddŵr Extech ExStik PH100 ​​a PH110
Comprehensive user manual for the Extech ExStik Waterproof pH Meters, models PH100 and PH110. Provides detailed instructions on operation, calibration, electrode maintenance, battery replacement, safety precautions, specifications, and troubleshooting for these high-accuracy pH testing devices.
Cynview Llawlyfr Defnyddiwr Mesurydd Dargludedd, TDS, Halenedd, Tymheredd Extech ExStik II EC400
Comprehensive user manual for the Extech ExStik II EC400 meter, detailing its features for measuring conductivity, total dissolved solids (TDS), salinity, and temperature. Includes setup, operation, calibration, maintenance, troubleshooting, and specifications.
Cynview Extech Comprehensive Catalog: Environmental Meters, Anemometers, and Test Equipment
Explore Extech's extensive range of high-quality environmental meters, thermo-anemometers, gas detectors, water quality testers, sound meters, light meters, and other precision test and measurement instruments designed for industrial, HVAC/R, and professional applications.
Cynview Extech PH220 Waterproof Palm pH Meter User Manual
User manual for the Extech PH220 Waterproof Palm pH Meter. Covers introduction, meter description, operation, calibration, measurement, storage, battery replacement, maintenance, troubleshooting, and specifications.
Cynview Llawlyfr Defnyddiwr Mesurydd Lleithder, Tymheredd, Llif Aer a Golau EXTECH 45170 4-mewn-1
Llawlyfr defnyddiwr ar gyfer yr EXTECH 45170, mesurydd 4-mewn-1 sy'n mesur lleithder, tymheredd, llif aer, a golau. Yn cynnwys gwybodaeth am weithrediad, manylebau, a gwarant.