U.S. Solid USS-DBS64-500

U.S. Solid 500g x 0.001g Lab Analytical Balance Instruction Manual

Model: USS-DBS64-500

1. Rhagymadrodd

This manual provides essential instructions for the proper setup, operation, maintenance, and troubleshooting of your U.S. Solid 500g x 0.001g Lab Analytical Balance. Designed for high-precision weighing, this balance offers reliable results with 0.001g (1mg) accuracy and a maximum capacity of 500g. It features multiple unit conversions, parts counting, percentage weighing, and manual calibration capabilities.

Please read this manual thoroughly before using the balance to ensure accurate measurements and safe operation.

2. Rhagofalon Diogelwch

  • Always operate the balance on a stable, level, and vibration-free surface.
  • Avoid placing the balance near strong electromagnetic fields, direct sunlight, or rapid temperature changes.
  • Peidiwch â gorlwytho'r cydbwysedd y tu hwnt i'w gapasiti uchaf o 500g.
  • Keep the balance and its components clean and free from dust and corrosive substances.
  • Use only the provided power adapter or a compatible 9V battery.
  • Do not attempt to disassemble or repair the balance yourself. Contact qualified service personnel if needed.

3. Beth sydd yn y Bocs

Carefully unpack all components and check against the following list:

  • Scale Body x1
  • Scale Pan x1
  • Cord pwer x1
  • Calibration Weight x1
  • Llawlyfr Cyfarwyddiadau x1

4. Cynnyrch Drosview

Familiarize yourself with the main components of your analytical balance.

Blaen view of the U.S. Solid Lab Analytical Balance with windshield and display showing 0.000g.
Ffigur 4.1: Blaen view of the balance. Shows the main display, control buttons, weighing pan, and windshield.
Ochr view of the U.S. Solid Lab Analytical Balance, highlighting the compact design.
Ffigur 4.2: Ochr view of the balance, illustrating its compact form factor.
Cefn view of the U.S. Solid Lab Analytical Balance, showing the power input.
Ffigur 4.3: Cefn view of the balance, including the power input port.
Close-up of the bubble level on the U.S. Solid Lab Analytical Balance for horizontal adjustment.
Ffigur 4.4: Detail of the integrated bubble level for precise horizontal adjustment.
Ffrwydrodd view of the U.S. Solid Lab Analytical Balance components with dimensions.
Ffigur 4.5: Ffrwydrodd view showing all components and their dimensions.

5. Gosod

5.1 Dadbacio ac Archwiliad Cychwynnol

Carefully remove the balance from its packaging. Inspect all components for any signs of damage. If any damage is found, do not use the balance and contact customer support immediately.

5.2 Lleoliad

Place the balance on a firm, level, and stable surface. Ensure the location is free from vibrations, drafts, direct sunlight, and rapid temperature fluctuations, which can affect measurement accuracy.

5.3 Lefelu'r Cydbwysedd

The balance must be perfectly level for accurate readings. Use the adjustable feet at the bottom of the balance to center the bubble in the integrated level indicator (refer to Figure 4.4).

Image showing the horizontal adjustment process with the bubble level.
Ffigur 5.1: Horizontal Adjustment. Adjust leveling feet until the bubble is centered. Proper leveling is essential for accuracy.

5.4 Cysylltiad Pwer

Connect the provided power cord to the balance and a suitable power outlet (100-240V). Alternatively, the balance can be powered by a 9V battery for portable use.

5.5 Warm-up Time

After connecting power, allow the balance to warm up for at least 30 minutes before taking measurements. This stabilizes the internal components and ensures optimal accuracy.

6. Cyfarwyddiadau Gweithredu

6.1 Pŵer Ymlaen/Diffodd

  • Gwasgwch y YMLAEN / I FFWRDD button to turn the balance on.
  • Pwyswch a dal y YMLAEN / I FFWRDD button to turn the balance off.

6.2 Pwyso Sylfaenol

To perform a basic weight measurement:

  1. Ensure the balance is level and has completed its warm-up.
  2. Gwasgwch y TARE / ZERO botwm i osod yr arddangosfa i 0.000g.
  3. Carefully place the item to be weighed on the center of the weighing pan.
  4. Arhoswch i'r darlleniad sefydlogi. Bydd y pwysau sefydlog yn cael ei arddangos.

Fideo 6.1: Demonstration of basic weighing, unit conversion, and percentage function on the U.S. Solid Lab Analytical Balance.

6.3 Swyddogaeth Tare

The tare function allows you to subtract the weight of a container, so only the net weight of the contents is measured.

  1. Rhowch gynhwysydd gwag ar y badell bwyso.
  2. Gwasgwch y TARE / ZERO botwm. Bydd yr arddangosfa yn ailosod i 0.000g.
  3. Add the material to be weighed into the container. The display will show the net weight of the material.

6.4 Trosi Uned

The balance supports multiple weighing units. To change the unit:

  • Gwasgwch y UNED button repeatedly to cycle through available units: g (grams), ct (carats), oz (ounces), ozt (troy ounces), lb (pounds), gn (grains), t (metric tons), dwt (pennyweights), kg (kilograms), dr (drams), tlT (Taiwan Tael), N (Newtons), and % (percentaga).
  • The selected unit will be displayed on the screen.
Image showing the various measuring units available on the balance.
Ffigur 6.1: Available Measuring Units and Piece Counting options.

6.5 Parts Counting (PCS)

This function allows you to count a large number of identical items by weight.

  1. Place a known quantity of items (e.g., 10, 20, 50, 100, 150, 200, 250, 500) on the pan.
  2. Gwasgwch y PCS botwm.
  3. Select the corresponding sample quantity using the UNED botwm.
  4. Gwasgwch PCS again to confirm. The balance is now in counting mode.
  5. Add more items, and the display will show the total count.

6.6 Canrantage Pwyso

Y percentage function is useful for formulation and quality control, allowing you to measure a sample's weight as a percentage o bwysau cyfeirio.

  1. Rhowch y cyfeirnod sample ar y badell bwyso.
  2. Gwasgwch y UNED button until '%' is displayed. The reference weight will be set as 100%.
  3. Tynnwch y cyfeirnod sample.
  4. Place subsequent samples on the pan. Their weight will be displayed as a percentage o'r pwysau cyfeirio.

6.7 Graddnodi

Regular calibration ensures the accuracy of your balance. It is recommended to calibrate the balance upon initial setup, after relocation, or if you suspect inaccurate readings.

  1. Ensure the weighing pan is empty and the balance is stable.
  2. Pwyswch a dal y CAL button until 'CAL' appears on the display.
  3. The display will then show the required calibration weight (e.g., 500.000g).
  4. Carefully place the specified calibration weight (provided) onto the center of the weighing pan.
  5. Wait for the balance to automatically calibrate. Once complete, the display will show 'PASS' or return to normal weighing mode.
  6. Remove the calibration weight.
Image showing the control panel buttons including ON/OFF, UNIT, PCS, CAL, and TARE/ZERO.
Ffigur 6.2: Control Panel Buttons. Key functions include ON/OFF, UNIT, PCS (Parts Counting), CAL (Calibration), and TARE/ZERO.

7. cynnal

7.1 Glanhau

To maintain accuracy and prolong the life of your balance:

  • Datgysylltwch y pŵer bob amser cyn glanhau.
  • Wipe the exterior of the balance with a soft, damp lliain. Defnyddiwch lanedydd ysgafn os oes angen.
  • Do not use abrasive cleaners, solvents, or immerse the balance in water.
  • Clean the stainless steel weighing pan and the windshield regularly to prevent accumulation of dust or residues.

7.2 Storio

When not in use, store the balance in a clean, dry environment, away from extreme temperatures and humidity. Keep the windshield in place to protect the weighing mechanism.

8. Datrys Problemau

This section addresses common issues you might encounter with your analytical balance.

ProblemAchos PosiblAteb
Darlleniadau anghywir neu amrywiol
  • Balance not level
  • Vibrations or drafts
  • Temperature changes
  • Needs calibration
  • Gorlwytho
  • Adjust leveling feet (Figure 5.1)
  • Relocate balance to a stable environment
  • Allow balance to acclimate to room temperature
  • Perform calibration (Section 6.7)
  • Ensure weight is within 500g capacity
Display shows 'Err' or other error codesInternal fault or overload condition.
  • Remove all items from the pan.
  • Turn off the balance, wait 30 seconds, then turn it back on.
  • Os yw'r gwall yn parhau, cysylltwch â chymorth cwsmeriaid.
Balance does not power on
  • Dim cyflenwad pŵer
  • Faulty power adapter/battery
  • Gwiriwch gysylltiad y llinyn pŵer a'r allfa.
  • Replace 9V battery if using battery power.
  • Test power adapter with another device if possible.
Cannot detect small weight changes (e.g., 10-15mg)
  • Environmental interference (drafts, vibrations)
  • Balance not properly calibrated
  • Warm-up time insufficient
  • Ensure stable environment, use windshield.
  • Perform calibration (Section 6.7).
  • Allow sufficient warm-up time (Section 5.5).

9. manylebau

Technical specifications for the U.S. Solid 500g x 0.001g Lab Analytical Balance.

NodweddManylyn
BrandUnol Daleithiau solet
Rhif ModelUSS-DBS64-500
Terfyn Pwysau500 gram
Cywirdeb Darllen Allan0.001 Grams (1mg)
Math ArddangosLCD (Goleuadau ôl)
Pwyso Graddfa MathGraddfa Balans
Unedau Mesurg, ct, oz, ozt, lb, gn, t, dwt, kg, dr, tlT, N, %
Cyflenwad PŵerAC (100-240V) / DC (9V Battery)
Math o DdeunyddAcrylonitrile Butadiene Styrene (ABS), Stainless Steel
Pwysau Eitem3.64 Bunt
Cydrannau wedi'u CynnwysScale Body, Scale Pan, Power Cord, Calibration Weight, Manual

10. Gwarant a Chefnogaeth

U.S. Solid products are manufactured to high-quality standards. For information regarding product warranty, please refer to the warranty card included with your purchase or visit the official U.S. Solid websafle.

For technical support, troubleshooting assistance, or to inquire about replacement parts, please contact U.S. Solid customer service. Contact details can typically be found on the product packaging or the official brand websafle.