1. Rhagymadrodd
The Hantek DSO5000P series digital storage oscilloscopes (DSO) are versatile and high-performance instruments designed for various electronic measurement tasks. This series includes models DSO5072P, DSO5102P, and DSO5202P, offering bandwidths of 70MHz, 100MHz, and 200MHz respectively, with a 1GSa/s real-time sample rate. These oscilloscopes feature a large 7.0-inch color display (WVGA 800x480) and a record length of up to 40K, making them suitable for detailed signal analysis in educational, industrial, and research environments.
This manual provides essential information for the safe and efficient operation, setup, and maintenance of your Hantek DSO5000P series oscilloscope.
2. Nodweddion Allweddol
- Bandwidths: 70MHz (DSO5072P), 100MHz (DSO5102P), 200MHz (DSO5202P)
- 1GSa/s Real-Time Sample Cyfradd
- Large 7.0-inch WVGA (800x480) Color Display
- Record Length up to 40K points
- Multiple Automatic Measurements
- Trigger Modes: Edge, Pulse Width, Line Selectable Video, Slope, Overtime
- USB Host and Device Connectivity for PC Analysis
- Four Math Functions, including FFT
3. Gwybodaeth Diogelwch
Always observe the following safety precautions to prevent injury and avoid damage to the instrument or connected devices.
- Ffynhonnell Pwer: Use only the specified power adapter and voltage. Ensure the power cord is properly grounded.
- Amgylchedd: Operate the oscilloscope in a dry, well-ventilated area, away from direct sunlight, dust, and corrosive gases. Avoid extreme temperatures.
- Archwilwyr: Use only probes rated for the voltage and current being measured. Ensure probes are properly connected before applying power.
- Awyru: Do not block the ventilation openings on the instrument.
- Gwasanaethu: Refer all servicing to qualified service personnel. Do not attempt to open or repair the instrument yourself.
4. Cynnwys Pecyn
Gwiriwch fod eich pecyn yn cynnwys yr eitemau canlynol:
- Hantek DSO5000P Series Digital Storage Oscilloscope
- Two Passive Probes (1X/10X switchable)
- Cord Pŵer
- Cebl USB
- CD-ROM with PC Software and User Manual
- Canllaw Cyflym
5. Gosod
Follow these steps to set up your oscilloscope for initial use.
5.1 Dros Dro Corfforolview
Familiarize yourself with the front, rear, and side panels of the oscilloscope.

This image displays the front, rear, and side views of the Hantek DSO5102P oscilloscope, highlighting its physical layout and connection ports.
5.2 Cysylltiad Pwer
- Connect the power cord to the AC input on the rear panel of the oscilloscope.
- Plygiwch ben arall y llinyn pŵer i mewn i allfa pŵer AC wedi'i seilio.
5.3 Cysylltiad a Digollediad y Prob
- Connect a passive probe to one of the input channels (CH1 or CH2) on the front panel.
- Attach the probe ground clip to the ground terminal on the oscilloscope.
- Set the probe attenuation switch (if applicable) to 10X.
- Connect the probe tip to the probe compensation output terminal (usually a square wave signal).
- Power on the oscilloscope by pressing the power button.
- Adjust the probe compensation trimmer on the probe until the square wave displayed on the screen has flat tops and bottoms, without overshoot or undershoot.

This image shows the Hantek DSO5102P oscilloscope actively measuring signals on a circuit board using a connected probe.
6. Cyfarwyddiadau Gweithredu
This section covers basic operation of the DSO5000P series oscilloscopes.
6.1 Rheolaethau Panel Blaen
The front panel is organized into several functional areas:
- Rheolyddion Fertigol: Addasu cyftage per division (Volts/Div) and position for each channel.
- Rheolyddion Llorweddol: Adjust time per division (Sec/Div) and horizontal position.
- Rheolyddion Sbardun: Set trigger level, mode, and type to stabilize waveforms.
- Botymau Swyddogaeth (F1-F6): Context-sensitive buttons for menu navigation and selections.
- Measurement Buttons: Access automatic measurement functions.
- Utility Buttons: Save/Recall, Acquire, Display, Cursor, Math, etc.

This image provides a detailed front view of the Hantek DSO5102P oscilloscope, illustrating the layout of its display, control knobs, and function buttons.
6.2 Basic Waveform Display
- Connect a signal source to CH1.
- Gwasgwch y AUTOSET button for automatic waveform display optimization.
- Addaswch y SEFYLLFA FERTIGOL a GWIRFODDOLWYR / DIV knobs to fine-tune the vertical scale and position of the waveform.
- Addaswch y SEFYLLFA HORIZONTAL a SEC/DIV knobs to fine-tune the horizontal scale and position.
- Addaswch y TRIGGER LEVEL knob to stabilize the waveform.

This image displays several examples of waveforms captured and analyzed on the oscilloscope screen, demonstrating its measurement capabilities.

This image presents a grid of six different oscilloscope screen captures, illustrating the device's ability to display and analyze various signal types simultaneously.
6.3 Using Automatic Measurements
Gwasgwch y MESUR button to access the automatic measurement menu. Use the function buttons (F1-F6) to select desired measurements such as Vpp, Vmax, Vmin, Frequency, Period, Rise Time, Fall Time, etc. The results will be displayed on the screen.
6.4 PC Connectivity
Connect the oscilloscope to a computer using the provided USB cable. Install the PC software from the included CD-ROM. This software allows for real-time data analysis, waveform storage, and remote control of the oscilloscope from your computer.
7. cynnal
7.1 Glanhau
To clean the instrument, use a soft cloth dampened with mild detergent and water. Do not use abrasive cleaners or solvents that may damage the casing or screen. Ensure the instrument is powered off and disconnected from all power sources before cleaning.
7.2 Storio
When not in use, store the oscilloscope in a clean, dry environment, away from extreme temperatures and humidity. Protect the screen from scratches and impacts.
8. Datrys Problemau
If you encounter issues with your oscilloscope, refer to the following common problems and solutions:
- Dim Arddangosfa: Ensure the power cord is securely connected and the power button is pressed. Check the brightness settings.
- No Waveform: Verify that probes are correctly connected to the input channels and the signal source. Check probe compensation and attenuation settings. Use the AUTOSET function.
- Tonffurf Ansefydlog: Adjust the trigger level and trigger mode. Ensure the trigger source is correctly selected.
- Mesuriadau Anghywir: Check probe attenuation settings (1X/10X) and ensure they match the oscilloscope's channel settings. Perform probe compensation.
- PC Software Connection Issues: Ensure the USB cable is properly connected and the correct drivers are installed on your PC. Restart both the oscilloscope and the PC.
For persistent issues, contact Hantek customer support.
9. manylebau
Detailed technical specifications for the Hantek DSO5000P series oscilloscopes.

This image presents a comprehensive table detailing the technical specifications for the DSO5072P, DSO5102P, and DSO5202P models.
| Paramedr | Gwerth |
|---|---|
| Gwneuthurwr | Hantek |
| Rhif Rhan | Q-DSO5202P |
| Pwysau Eitem | 3 Cilogram |
| Dimensiynau Cynnyrch | 10 x 31.5 x 14 cm; 3 kg |
| Terfynwyd gan y Gwneuthurwr | Nac ydw |
| Size (Model) | DSO5202P |
| Lliw | Amryliw |
| Arddull | Bench Oscilloscope |
| Deunydd | Biwtadïen Acrylonitrile Styrene |
| Wattage | 1 Gw |
| System Fesur | Metrig |
| Cywirdeb Mesur | 0.3% |
| Ardystiadau | CE, ISO 9001, RoHS |
| Cydrannau wedi'u Cynnwys | DSO5202P (main unit, probes, power cord, USB cable, software CD) |
| Dimensiynau Eitem (LxWxH) | 10 x 31.5 x 14 cm |
| ASIN | B0C72HY47Z |
| Dyddiad Ar Gael Cyntaf ar Amazon.co.jp | 2023/6/4 |
| Paramedr | DSO5202P | DSO5102P | DSO5072P |
|---|---|---|---|
| Lled band | 200MHz | 100MHz | 70MHz |
| Sample Rate (Real-Time) | 1GSa/e | ||
| Equivalent Sample Cyfradd | 25GS/e | ||
| Moddau Caffael | Normal, Peak Detect, Average (4, 8, 16, 32, 64, 128) | ||
| Hyd y Cofnod | 40K | ||
| Inputs Coupling | AC, DC, GND | ||
| Rhwystriant Mewnbwn | 1MΩ ±2% || 20pF ±3pF | ||
| Gwanhau Profi | 1X, 10X, 100X, 1000X | ||
| Uchafswm Mewnbwn Voltage | CAT I and CAT II: 300VRMS (10×), Installation Category II: derate at 20dB/decade above 100kHz to 13V peak AC at 3MHz*. For non-sinusoidal waveforms, peak value must be less than 450V. Excursion above 300V should be of less than 100ms duration. RMS signal level including all DC components removed through AC coupling must be limited to 300V. If these values are exceeded, damage to the oscilloscope may occur. | ||
| Horizontal Sample Ystod Ardrethi | 500MS/s – 1GS/s | ||
| Rhyngosod Tonffurf | (pechod x)/x | ||
| SEC/DIV Range | 2ns/div to 80us/div | ||
| Sample Rate and Delay Time Accuracy | ±50ppm (at over any ≥1ms time interval) | ||
| Ystod Gwrthbwyso | 2ns/div to 8ns/div: 20ns/div to 80us/div: (-8div x s/div) to 40ms; 200us/div to 80s/div: (-8div x s/div) to 400s | ||
| Delta Time Measurement Accuracy (Full Bandwidth) | Single-shot, Normal mode: ± (1 sample interval + 100ppm × reading + 0.6ns); >16 averages: ± (1 sample interval + 100ppm × reading + 0.4ns); Sample interval = s/div ÷ 200 | ||
| Datrysiad Fertigol | 8-bit resolution, all channel sampled simultaneously | ||
| Amrediad Swydd | 2mV/div to 20V/div: ±400mV; 50mV/div to 20V/div: ±2V; 500mV/div to 2V/div: ±40V; 5V/div to 10V/div: ±250V | ||
| Rise Time at BNC (typical) | 1.8ns | 3.5ns | 5ns |
| Swyddogaethau Mathemateg | +, -, ×, ÷, FFT | ||
| FFT Windows | Hanning, Flatop, Rectangular, Bartlett, Blackman; 1024 samppwynt | ||
| Terfyn Lled Band | 20MHz | ||
| Low Frequency Response (-3dB) | ≤10Hz at BNC | ||
| Cywirdeb Ennill DC | ±3% for Normal or Average acquisition mode, 10V/div to 10mV/div; ±4% for Normal or Average acquisition mode, 5mV/div to 2mV/div | ||
| DC Measurement Accuracy | When vertical displacement is zero, and N ≥16: ± (3% × reading + 0.1div + 1mV) only 10mV/div or greater is selected. | ||
10. Gwarant a Chefnogaeth
Hantek products are designed for reliability and performance. This product is covered by a standard manufacturer's warranty against defects in materials and workmanship. Please refer to the warranty card included in your package for specific terms and conditions.
For technical support, troubleshooting assistance, or warranty claims, please contact Hantek customer service through their official website or the contact information provided with your product. When contacting support, please have your product model number and purchase date available.