Hantek DSO5072P, DSO5102P, DSO5202P

Hantek DSO5000P Series Digital Storage Oscilloscope User Manual

Models: DSO5072P, DSO5102P, DSO5202P

1. Rhagymadrodd

The Hantek DSO5000P series digital storage oscilloscopes (DSO) are versatile and high-performance instruments designed for various electronic measurement tasks. This series includes models DSO5072P, DSO5102P, and DSO5202P, offering bandwidths of 70MHz, 100MHz, and 200MHz respectively, with a 1GSa/s real-time sample rate. These oscilloscopes feature a large 7.0-inch color display (WVGA 800x480) and a record length of up to 40K, making them suitable for detailed signal analysis in educational, industrial, and research environments.

This manual provides essential information for the safe and efficient operation, setup, and maintenance of your Hantek DSO5000P series oscilloscope.

2. Nodweddion Allweddol

  • Bandwidths: 70MHz (DSO5072P), 100MHz (DSO5102P), 200MHz (DSO5202P)
  • 1GSa/s Real-Time Sample Cyfradd
  • Large 7.0-inch WVGA (800x480) Color Display
  • Record Length up to 40K points
  • Multiple Automatic Measurements
  • Trigger Modes: Edge, Pulse Width, Line Selectable Video, Slope, Overtime
  • USB Host and Device Connectivity for PC Analysis
  • Four Math Functions, including FFT

3. Gwybodaeth Diogelwch

Always observe the following safety precautions to prevent injury and avoid damage to the instrument or connected devices.

  • Ffynhonnell Pwer: Use only the specified power adapter and voltage. Ensure the power cord is properly grounded.
  • Amgylchedd: Operate the oscilloscope in a dry, well-ventilated area, away from direct sunlight, dust, and corrosive gases. Avoid extreme temperatures.
  • Archwilwyr: Use only probes rated for the voltage and current being measured. Ensure probes are properly connected before applying power.
  • Awyru: Do not block the ventilation openings on the instrument.
  • Gwasanaethu: Refer all servicing to qualified service personnel. Do not attempt to open or repair the instrument yourself.

4. Cynnwys Pecyn

Gwiriwch fod eich pecyn yn cynnwys yr eitemau canlynol:

  • Hantek DSO5000P Series Digital Storage Oscilloscope
  • Two Passive Probes (1X/10X switchable)
  • Cord Pŵer
  • Cebl USB
  • CD-ROM with PC Software and User Manual
  • Canllaw Cyflym

5. Gosod

Follow these steps to set up your oscilloscope for initial use.

5.1 Dros Dro Corfforolview

Familiarize yourself with the front, rear, and side panels of the oscilloscope.

Blaen, Cefn, ac Ochr views of the Hantek DSO5102P Digital Storage Oscilloscope

This image displays the front, rear, and side views of the Hantek DSO5102P oscilloscope, highlighting its physical layout and connection ports.

5.2 Cysylltiad Pwer

  1. Connect the power cord to the AC input on the rear panel of the oscilloscope.
  2. Plygiwch ben arall y llinyn pŵer i mewn i allfa pŵer AC wedi'i seilio.

5.3 Cysylltiad a Digollediad y Prob

  1. Connect a passive probe to one of the input channels (CH1 or CH2) on the front panel.
  2. Attach the probe ground clip to the ground terminal on the oscilloscope.
  3. Set the probe attenuation switch (if applicable) to 10X.
  4. Connect the probe tip to the probe compensation output terminal (usually a square wave signal).
  5. Power on the oscilloscope by pressing the power button.
  6. Adjust the probe compensation trimmer on the probe until the square wave displayed on the screen has flat tops and bottoms, without overshoot or undershoot.
Hantek DSO5102P oscilloscope in use with a probe connected to a circuit board

This image shows the Hantek DSO5102P oscilloscope actively measuring signals on a circuit board using a connected probe.

6. Cyfarwyddiadau Gweithredu

This section covers basic operation of the DSO5000P series oscilloscopes.

6.1 Rheolaethau Panel Blaen

The front panel is organized into several functional areas:

  • Rheolyddion Fertigol: Addasu cyftage per division (Volts/Div) and position for each channel.
  • Rheolyddion Llorweddol: Adjust time per division (Sec/Div) and horizontal position.
  • Rheolyddion Sbardun: Set trigger level, mode, and type to stabilize waveforms.
  • Botymau Swyddogaeth (F1-F6): Context-sensitive buttons for menu navigation and selections.
  • Measurement Buttons: Access automatic measurement functions.
  • Utility Buttons: Save/Recall, Acquire, Display, Cursor, Math, etc.
Blaen view of the Hantek DSO5102P Digital Storage Oscilloscope with labels for controls

This image provides a detailed front view of the Hantek DSO5102P oscilloscope, illustrating the layout of its display, control knobs, and function buttons.

6.2 Basic Waveform Display

  1. Connect a signal source to CH1.
  2. Gwasgwch y AUTOSET button for automatic waveform display optimization.
  3. Addaswch y SEFYLLFA FERTIGOL a GWIRFODDOLWYR / DIV knobs to fine-tune the vertical scale and position of the waveform.
  4. Addaswch y SEFYLLFA HORIZONTAL a SEC/DIV knobs to fine-tune the horizontal scale and position.
  5. Addaswch y TRIGGER LEVEL knob to stabilize the waveform.
Multiple oscilloscope screen captures showing various waveforms and measurement results

This image displays several examples of waveforms captured and analyzed on the oscilloscope screen, demonstrating its measurement capabilities.

Grid of six oscilloscope screen captures showing different signal types and analysis

This image presents a grid of six different oscilloscope screen captures, illustrating the device's ability to display and analyze various signal types simultaneously.

6.3 Using Automatic Measurements

Gwasgwch y MESUR button to access the automatic measurement menu. Use the function buttons (F1-F6) to select desired measurements such as Vpp, Vmax, Vmin, Frequency, Period, Rise Time, Fall Time, etc. The results will be displayed on the screen.

6.4 PC Connectivity

Connect the oscilloscope to a computer using the provided USB cable. Install the PC software from the included CD-ROM. This software allows for real-time data analysis, waveform storage, and remote control of the oscilloscope from your computer.

7. cynnal

7.1 Glanhau

To clean the instrument, use a soft cloth dampened with mild detergent and water. Do not use abrasive cleaners or solvents that may damage the casing or screen. Ensure the instrument is powered off and disconnected from all power sources before cleaning.

7.2 Storio

When not in use, store the oscilloscope in a clean, dry environment, away from extreme temperatures and humidity. Protect the screen from scratches and impacts.

8. Datrys Problemau

If you encounter issues with your oscilloscope, refer to the following common problems and solutions:

  • Dim Arddangosfa: Ensure the power cord is securely connected and the power button is pressed. Check the brightness settings.
  • No Waveform: Verify that probes are correctly connected to the input channels and the signal source. Check probe compensation and attenuation settings. Use the AUTOSET function.
  • Tonffurf Ansefydlog: Adjust the trigger level and trigger mode. Ensure the trigger source is correctly selected.
  • Mesuriadau Anghywir: Check probe attenuation settings (1X/10X) and ensure they match the oscilloscope's channel settings. Perform probe compensation.
  • PC Software Connection Issues: Ensure the USB cable is properly connected and the correct drivers are installed on your PC. Restart both the oscilloscope and the PC.

For persistent issues, contact Hantek customer support.

9. manylebau

Detailed technical specifications for the Hantek DSO5000P series oscilloscopes.

Detailed specifications table for Hantek DSO5072P, DSO5102P, and DSO5202P oscilloscopes

This image presents a comprehensive table detailing the technical specifications for the DSO5072P, DSO5102P, and DSO5202P models.

Manylebau Cyffredinol
ParamedrGwerth
GwneuthurwrHantek
Rhif RhanQ-DSO5202P
Pwysau Eitem3 Cilogram
Dimensiynau Cynnyrch10 x 31.5 x 14 cm; 3 kg
Terfynwyd gan y GwneuthurwrNac ydw
Size (Model)DSO5202P
LliwAmryliw
ArddullBench Oscilloscope
DeunyddBiwtadïen Acrylonitrile Styrene
Wattage1 Gw
System FesurMetrig
Cywirdeb Mesur0.3%
ArdystiadauCE, ISO 9001, RoHS
Cydrannau wedi'u CynnwysDSO5202P (main unit, probes, power cord, USB cable, software CD)
Dimensiynau Eitem (LxWxH)10 x 31.5 x 14 cm
ASINB0C72HY47Z
Dyddiad Ar Gael Cyntaf ar Amazon.co.jp2023/6/4
Manylebau Perfformiad
ParamedrDSO5202PDSO5102PDSO5072P
Lled band200MHz100MHz70MHz
Sample Rate (Real-Time)1GSa/e
Equivalent Sample Cyfradd25GS/e
Moddau CaffaelNormal, Peak Detect, Average (4, 8, 16, 32, 64, 128)
Hyd y Cofnod40K
Inputs CouplingAC, DC, GND
Rhwystriant Mewnbwn1MΩ ±2% || 20pF ±3pF
Gwanhau Profi1X, 10X, 100X, 1000X
Uchafswm Mewnbwn VoltageCAT I and CAT II: 300VRMS (10×), Installation Category II: derate at 20dB/decade above 100kHz to 13V peak AC at 3MHz*. For non-sinusoidal waveforms, peak value must be less than 450V. Excursion above 300V should be of less than 100ms duration. RMS signal level including all DC components removed through AC coupling must be limited to 300V. If these values are exceeded, damage to the oscilloscope may occur.
Horizontal Sample Ystod Ardrethi500MS/s – 1GS/s
Rhyngosod Tonffurf(pechod x)/x
SEC/DIV Range2ns/div to 80us/div
Sample Rate and Delay Time Accuracy±50ppm (at over any ≥1ms time interval)
Ystod Gwrthbwyso2ns/div to 8ns/div: 20ns/div to 80us/div: (-8div x s/div) to 40ms; 200us/div to 80s/div: (-8div x s/div) to 400s
Delta Time Measurement Accuracy (Full Bandwidth)Single-shot, Normal mode: ± (1 sample interval + 100ppm × reading + 0.6ns); >16 averages: ± (1 sample interval + 100ppm × reading + 0.4ns); Sample interval = s/div ÷ 200
Datrysiad Fertigol8-bit resolution, all channel sampled simultaneously
Amrediad Swydd2mV/div to 20V/div: ±400mV; 50mV/div to 20V/div: ±2V; 500mV/div to 2V/div: ±40V; 5V/div to 10V/div: ±250V
Rise Time at BNC (typical)1.8ns3.5ns5ns
Swyddogaethau Mathemateg+, -, ×, ÷, FFT
FFT WindowsHanning, Flatop, Rectangular, Bartlett, Blackman; 1024 samppwynt
Terfyn Lled Band20MHz
Low Frequency Response (-3dB)≤10Hz at BNC
Cywirdeb Ennill DC±3% for Normal or Average acquisition mode, 10V/div to 10mV/div; ±4% for Normal or Average acquisition mode, 5mV/div to 2mV/div
DC Measurement AccuracyWhen vertical displacement is zero, and N ≥16: ± (3% × reading + 0.1div + 1mV) only 10mV/div or greater is selected.

10. Gwarant a Chefnogaeth

Hantek products are designed for reliability and performance. This product is covered by a standard manufacturer's warranty against defects in materials and workmanship. Please refer to the warranty card included in your package for specific terms and conditions.

For technical support, troubleshooting assistance, or warranty claims, please contact Hantek customer service through their official website or the contact information provided with your product. When contacting support, please have your product model number and purchase date available.

© 2023 Hantek. All rights reserved.