Sartorius Cubis II Semi-Micro Balance MCA125S-2S00-I User Manual

Model: MCA125S-2S00-I

1. Rhagymadrodd

This manual provides essential instructions for the safe and efficient operation, setup, and maintenance of your Sartorius Cubis II Semi-Micro Balance, model MCA125S-2S00-I. This high-performance laboratory balance is designed for precise weighing applications requiring a 0.01 mg division and a maximum load of 120g. It features a modular design, an MCA Interface, and an integrated Draft Shield I to ensure accurate and repeatable results in R&D and analytical laboratories.

Sartorius Cubis II Semi-Micro Balance MCA125S-2S00-I

Figure 1: Sartorius Cubis II Semi-Micro Balance MCA125S-2S00-I with Draft Shield I. This image shows the complete balance unit, including the weighing pan and the protective draft shield.

2. Gwybodaeth Diogelwch

Always observe the following safety precautions to prevent injury and damage to the balance:

  • Read this entire manual before operating the balance.
  • Operate the balance only in a stable, level environment free from vibrations, drafts, and extreme temperature fluctuations.
  • Ensure the power supply matches the balance's requirements (100-240V).
  • Do not expose the balance to corrosive chemicals or excessive moisture.
  • Only use Sartorius-approved accessories and spare parts.
  • Datgysylltwch y cebl pŵer cyn glanhau neu wneud unrhyw waith cynnal a chadw.
  • Peidiwch â cheisio agor na thrwsio'r clorian eich hun. Cyfeiriwch y gwaith cynnal a chadw at bersonél cymwys.

3. Dadbacio ac Arolygiad Cychwynnol

Carefully unpack all components and check for any visible damage. Retain the original packaging for future transport or storage.

  1. Open the shipping carton and remove the balance unit and accessories.
  2. Gwiriwch fod yr holl gydrannau a restrir yn y rhestr bacio yn bresennol.
  3. Inspect the balance for any signs of damage incurred during shipping. If damage is found, contact your supplier immediately.

4. Gosod

4.1 Lleoliad

Place the balance on a sturdy, level, and vibration-free surface. Avoid locations near air conditioning vents, open windows, or direct sunlight, which can cause temperature fluctuations and affect weighing accuracy.

4.2 Draft Shield Assembly

The Cubis II MCA125S-2S00-I comes with Draft Shield I. Assemble the draft shield components according to the diagrams provided in the separate assembly guide (if applicable) or by carefully following the intuitive design. Ensure all glass panels are securely seated to prevent air currents from affecting measurements.

Draft Shield I Assembly

Figure 2: Illustrative image of the balance with Draft Shield I. This image highlights the transparent draft shield enclosure, crucial for protecting the weighing pan from air currents.

4.3 Lefelu'r Cydbwysedd

The balance must be perfectly level for accurate measurements. Use the leveling feet located at the base of the balance and the integrated spirit level (bubble level) to adjust until the bubble is centered.

  1. Locate the spirit level, usually at the front or side of the balance.
  2. Adjust the two rear leveling feet until the bubble is centered in the spirit level.
  3. Adjust the front leveling foot (if present) to fine-tune the level.

4.4 Cysylltiad Pwer

Connect the power adapter to the balance and then plug it into a suitable electrical outlet (100-240V AC). Allow the balance to warm up for at least 30 minutes after connecting power before performing any measurements to ensure optimal stability.

5. Cyfarwyddiadau Gweithredu

5.1 Troi Ymlaen/Diffodd

Gwasgwch y YMLAEN / I FFWRDD button (often indicated by a circle with a vertical line) to power on the balance. To turn off, press and hold the same button until the display shuts down.

5.2 Gweithdrefn Pwyso Sylfaenol

  1. Gwnewch yn siŵr bod y badell bwyso yn lân ac yn wag.
  2. Caewch y drysau sgrin drafft.
  3. Gwasgwch y TARE or ZERO botwm i osod yr arddangosfa i sero.
  4. Rhowch yr eitem i'w phwyso yn ofalus yng nghanol y badell bwyso.
  5. Caewch y drysau sgrin drafft.
  6. Wait for the stability indicator (usually a circle or asterisk) to appear on the display, indicating a stable reading.
  7. Read the weight displayed in milligrams (mg).
  8. Remove the item from the weighing pan.

5.3 Using the Tare Function

The tare function allows you to weigh the net weight of a substance by subtracting the weight of its container.

  1. Rhowch y cynhwysydd gwag ar y badell bwyso.
  2. Caewch y drysau sgrin drafft.
  3. Gwasgwch y TARE button. The display will show zero.
  4. Ychwanegwch y sylwedd i'w bwyso i'r cynhwysydd.
  5. Close the draft shield doors and wait for a stable reading. The displayed value is the net weight of the substance.

5.4 MCA Interface

The MCA Interface provides advanced control and data management capabilities. Refer to the separate MCA Interface user guide for detailed instructions on navigating menus, configuring settings, and connecting to external devices or software. This interface allows for customization of weighing applications, data logging, and compliance features.

6. graddnodi

Regular calibration is crucial for maintaining the accuracy of your balance. The Cubis II balance may feature internal calibration (isoCAL) or require external calibration using certified weights.

6.1 Internal Calibration (isoCAL)

If your balance is equipped with isoCAL, it will perform automatic internal calibration at set intervals or when temperature changes occur. You can also initiate it manually via the menu. Consult the MCA Interface user guide for specific steps.

6.2 Calibradu Allanol

For external calibration, you will need certified calibration weights appropriate for the balance's capacity and readability. The procedure typically involves:

  1. Accessing the calibration menu on the MCA Interface.
  2. Following the on-screen prompts to place and remove the specified calibration weights.
  3. Confirming the calibration process.

Always use certified weights and ensure they are clean and handled with gloves to avoid contamination.

7. cynnal

Proper maintenance ensures the longevity and accuracy of your balance.

7.1 Glanhau

  • Datgysylltwch y cydbwysedd o'r cyflenwad pŵer bob amser cyn glanhau.
  • Defnyddio lliain meddal, di-lint dampened with a mild cleaning agent (e.g., isopropyl alcohol or a mild soap solution) to clean the exterior surfaces.
  • Remove the weighing pan and draft shield components for separate cleaning. Wash them with mild detergent and rinse thoroughly. Ensure they are completely dry before reassembling.
  • Do not use abrasive cleaners, strong solvents, or immerse the balance in water.
  • Avoid spraying cleaning agents directly onto the balance; apply them to the cloth first.

7.2 Amodau Amgylcheddol

Maintain a stable environment. Avoid rapid temperature changes, high humidity, and strong air currents. Store the balance in a clean, dry place when not in use.

8. Datrys Problemau

This section addresses common issues you might encounter. For problems not listed here, contact Sartorius customer support.

ProblemAchos PosiblAteb
Darlleniadau ansefydlogDrafts, vibrations, unstable surface, temperature fluctuations, dirty weighing pan.Ensure draft shield is closed. Relocate balance to a stable, vibration-free area. Check leveling. Allow balance to acclimatize. Clean weighing pan.
Nid yw'r cydbwysedd yn troi ymlaenNo power supply, faulty power cable/adapter.Check power cable connection. Verify power outlet. Try a different power adapter if available.
Darlleniadau anghywirNot calibrated, incorrect leveling, temperature drift, overloading.Perform calibration. Re-level the balance. Allow warm-up time. Do not exceed 120g max load.
Arddangos neges gwallInternal fault, overload, communication error.Refer to the MCA Interface user guide for specific error codes. Turn off and restart the balance. If persistent, contact support.

9. manylebau

Key technical specifications for the Sartorius Cubis II Semi-Micro Balance MCA125S-2S00-I:

  • Model: MCA125S-2S00-I
  • Brand: Sartorius
  • Maximum Load (Maxload): 120g
  • Readout Accuracy (Division): 0.01 mg
  • Rhyngwyneb: MCA Interface
  • Tarian Drafft: Draft Shield I
  • Cyflenwad Pŵer: 100-240V
  • Cydrannau wedi'u Cynnwys: Cydbwysedd
  • Ffactor Ffurflen: Mini

10. Gwarant a Chefnogaeth

For detailed warranty information, please refer to the warranty card included with your product or visit the official Sartorius website. In case of technical issues, service requirements, or questions not covered in this manual, please contact your local Sartorius representative or authorized service center. Provide your model number (MCA125S-2S00-I) and serial number when requesting support.

Sartorius Contact Information: Cyfeiriwch at www.sartorius.com/en/global-contacts am fanylion cyswllt rhanbarthol.

© 2023 Sartorius. All rights reserved. Information subject to change without notice.